Correlation Study of Arterial Blood Pressure Level to the Amplitude of the Pressure Pulse Waveform

dc.contributor.authorVachtsevanos, G.Jen
dc.contributor.authorKalaitzakis K.Cen
dc.contributor.authorPapamarkos, Nen
dc.contributor.authorZiakas G.en
dc.contributor.authorEconomou K.en
dc.contributor.authorGemitzis K.en
dc.date.accessioned2024-10-31T16:09:49Z
dc.date.available2024-10-31T16:09:49Z
dc.date.issued1984
dc.date.submitted2015-09-23
dc.descriptionΔημοσίευση σε διεθνές περιοδικόel
dc.description.abstractAutomated blood pressure measurements are usually characterized by poor operational reliability and a considerable degree of complexity in performing the measurement. This paper introduces a new technique for the indirect measurement of the systolic and diastolic blood pressure of an individual. The technique is based upon a statistically consistent relationship between the amplitude of the pulsative pressure waveform at the systolic and diastolic points, and the amplitude of pulse signals detected when the artery is fully occluded. Clinical testing and statistical analysis techniques are used to derive appropriate numerical values for these relationships. The proposed procedure thus incorporates an adaptive measurement philosophy achieving minimum observer involvement and consequently high instrument accuracy. Overall measurement errors are maintained well within proposed standards for automated sphygmomanometers.en
dc.description.journalnumber6
dc.description.journalvolume1
dc.description.pagerange33-39
dc.format.extent7en
dc.identifierhttp://www.tuc.gr/fileadmin/users_data/elci/Kalaitzakis/J.02.pdf
dc.identifier10.1016/0141-5425(84)90007-4
dc.identifier.citationG.J. Vachtsevanos, K.C. Kalaitzakis, N.E. Papamarkos, G. Ziakas, K. Economou, and K. Gemitzis, "Correlation Study of Arterial Blood Pressure Level to the Amplitude of the Pressure Pulse Waveform," Jour. of Biomedical Engineering, vol. 6, no. 1, pp. 33-39, Jan. 1984. doi:10.1016/0141-5425(84)90007-4en
dc.identifier.urihttps://dspace.library.tuc.gr/handle/123456789/1011
dc.language.isoen
dc.publisherButterworth & Co Ltden
dc.relation.isreferencedbyBiomedical Engineeringen
dc.relation.replaces11651
dc.rightshttp://creativecommons.org/licenses/by/4.0/en
dc.subjectBlood pressureen
dc.subjectMeasurementen
dc.subjectClinical dataen
dc.subjectStatistical analysisen
dc.subjectErroren
dc.titleCorrelation Study of Arterial Blood Pressure Level to the Amplitude of the Pressure Pulse Waveformen
dc.typePeer-Reviewed Journal Publicationen
dc.typeΔημοσίευση σε Περιοδικό με Κριτέςel
dspace.entity.typePublication

Αρχεία