Design and development of a new electronic sphygmomanometer

Δεν υπάρχει διαθέσιμη μικρογραφία

Ημερομηνία

1985

Συγγραφείς

Vachtsevanos, G.J
Kalaitzakis K.C
Papamarkos, N

Τίτλος Εφημερίδας

Περιοδικό ISSN

Τίτλος τόμου

Εκδότης

Kluwer Academic Publishers

Περίληψη

The paper introduces a new technique for the indirect measurement of the systolic and diastolic blood pressure in humans. The technique is based upon a statistically consistent relationship between the amplitude of the pulsatile pressure waveform at the systolic and diastolic points and the amplitude of pulse signals detected when the artery is fully occluded. An adaptive measurement philosophy has been implemented in the design of an electronic sphygmomanometer which, in addition to a pressure transducer, contains suitable electronic instrumentation for processing and displaying the electronic signals. Verification of overall system accuracy is accomplished with direct comparison with manual auscultatory measurements. Clinical testing of a prototype indicates a satisfactory performance; measurement errors are maintained well within proposed standards for automated sphygmomanometers.

Περιγραφή

Δημοσίευση σε επιστημονικό περιοδικό

Λέξεις-κλειδιά

Electronic design, Electronic sphygmomanometer, Noninvasive blood pressure measurement

Παραπομπή

G.J. Vachtsevanos, K.C. Kalaitzakis, and N.E, Papamarkos, "Design and development of a new electronic sphygmomanometer," Medical and Biological Engineering and Computing ., vol. 23, no. 5, pp. 453-458, Sep. 1985. doi:10.1007/BF02448933

Έχει διάδοχο το τεκμήριο

Είναι διάδοχο του τεκμηρίου

Περιέχει το τεκμήριο

Είναι μέρος του τεκμηρίου

Αναφέρει το τεκμήριο

Αναφέρεται από το τεκμήριο

Έπεται το τεκμήριο

Προηγείται του τεκμηρίου

Έχει ως έκδοση το τεκμήριο

Αποτελεί έκδοση του τεκμηρίου

Έχει ως συμπληρωματικό το τεκμήριο

Είναι συμπληρωματικό του τεκμηρίου

Έχει μετατραπει στο τεκμήριο

Αποτελεί μετατροπή του τεκμηρίου