Design and development of a new electronic sphygmomanometer

dc.contributor.authorVachtsevanos, G.Jen
dc.contributor.authorKalaitzakis K.Cen
dc.contributor.authorPapamarkos, Nen
dc.date.accessioned2024-10-31T15:33:46Z
dc.date.available2024-10-31T15:33:46Z
dc.date.issued1985
dc.date.submitted2015-09-25
dc.descriptionΔημοσίευση σε επιστημονικό περιοδικόel
dc.description.abstractThe paper introduces a new technique for the indirect measurement of the systolic and diastolic blood pressure in humans. The technique is based upon a statistically consistent relationship between the amplitude of the pulsatile pressure waveform at the systolic and diastolic points and the amplitude of pulse signals detected when the artery is fully occluded. An adaptive measurement philosophy has been implemented in the design of an electronic sphygmomanometer which, in addition to a pressure transducer, contains suitable electronic instrumentation for processing and displaying the electronic signals. Verification of overall system accuracy is accomplished with direct comparison with manual auscultatory measurements. Clinical testing of a prototype indicates a satisfactory performance; measurement errors are maintained well within proposed standards for automated sphygmomanometers.en
dc.description.journalnumber23
dc.description.journalvolume5
dc.description.pagerange453-458
dc.format.extent6en
dc.identifierhttp://www.tuc.gr/fileadmin/users_data/elci/Kalaitzakis/J.06.pdf
dc.identifier10.1007/BF02448933
dc.identifier.citationG.J. Vachtsevanos, K.C. Kalaitzakis, and N.E, Papamarkos, "Design and development of a new electronic sphygmomanometer," Medical and Biological Engineering and Computing ., vol. 23, no. 5, pp. 453-458, Sep. 1985. doi:10.1007/BF02448933en
dc.identifier.urihttps://dspace.library.tuc.gr/handle/123456789/618
dc.language.isoen
dc.publisherKluwer Academic Publishersen
dc.relation.isreferencedbyMedical and Biological Engineering and Computingen
dc.relation.replaces11945
dc.rightshttp://creativecommons.org/licenses/by/4.0/en
dc.subjectElectronic designen
dc.subjectElectronic sphygmomanometeren
dc.subjectNoninvasive blood pressure measurementen
dc.titleDesign and development of a new electronic sphygmomanometeren
dc.typePeer-Reviewed Journal Publicationen